Operation
For accurate results the CO meter should be used at room
temperature. If the instrument has been stored in cool or hot
conditions then allow time to reach room temperature prior to use
Install the PP3 battery by sliding open the battery cover, clipping the
battery in place and replacing the cover. Insert the mouthpiece
adapter into the MicroCO meter and then insert a disposable
mouthpiece into the adapter. Turn the unit on by selecting the
upper or middle position on the central slide switch and the display
will show the software version number.
This will appear momentarily whilst correction for ambient levels is
executed. During this time the unit must not be exposed to raised
concentrations of CO. The buzzer will then sound and the display
will change to:
Upon hearing the buzzer instruct the subject to inspire fully and hold
their breath for 20 seconds. The display will count down from 20 to
0 as an aid to timing the breath holding period. The green indicator
will then illuminate and the display will change to:
BLOW