CareFusion Smoke Check Meter Especificaciones Pagina 31

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 30
28
Symbols
Type B device
Environment
This instrument complies with directive EN60601-1-2 electromagnetic
compatibility but can be affected by cellular phones and by
electromagnetic interference exceeding levels specified in EN 50082-
1:1992
Specifications
Sensor type Electro-chemical fuel cell
Range 0 - 20 ppm
Detected levels: Display:
0 to 6ppm 0-6 ppm with green indicator
7 to 10ppm 7-10 ppm with amber indicator
11 to 20ppm 11-20 ppm with red indicator
>20 ppm 20+ ppm with red indicator + alarm
Accuracy +/-5% of reading or 1ppm whichever
is the greater
Sensitivity drift 0.5%/°C
Sensor life 2 to 5 years
Response time < 15 sec (to 90% of reading)
Hydrogen cross sensitivity <15%
Operating temperature 15 – 25 °C
Operating pressure Atmospheric +/- 10%
Pressure coefficient 0.02% signal per mBar
Relative humidity 15 - 90% continuous
(Non condensing) (0 - 99% intermittent)
Baseline drift 0ppm (auto-zero)
Long term drift < 2% signal loss per month
Vista de pagina 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 63 64

Comentarios a estos manuales

Sin comentarios