CareFusion Smoke Check Meter Especificaciones Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
16
Environment
This instrument complies with directive EN60601-1-2
electromagnetic compatibility but can be affected by cellular phones
and by electromagnetic interference exceeding levels specified in EN
50082-1:1992
Specifications
Sensor type Electro-chemical fuel cell
Range 0 - 100 ppm
Resolution 1 ppm
Green indicator light 0 to 6ppm
Amber indicator light 7 to 10ppm
Red indicator light 11 to 20ppm
Flashing red light 20+
Accuracy +/-5% of reading or 1ppm whichever
is the greater
Sensitivity drift 0.5%/°C
Sensor life 2 to 5 years
Response time < 15 sec (to 90% of reading)
Hydrogen cross sensitivity <15%
Operating temperature 15 - 25 °C
Operating pressure Atmospheric +/- 10%
Pressure coefficient 0.02% signal per mBar
Relative humidity 15 - 90% continuous
(Non condensing) (0 - 99% intermittent)
Baseline drift 0ppm (auto-zero)
Long term drift < 2% signal loss per month
Power source Single Alkaline 9 volt PP3
Main battery life 30 hours of continuous use equivalent
to approximately 2000 tests
Internal battery life 10 years
Weight 180 g (Including battery)
Vista de pagina 17