60
carefusion.com
Filters, HCHs & HMEs
003020
Air
Life
™
Nonfiltered BHME
003021
Air
Life
™
Filtered BHME
Tidal Volumes 250 to 1,250 mL 250 to 1,250 mL
Weight 43 g 44 g
Dead Space 68 mL 68 mL
Moisture Loss
250 mL = 9.1 mg H2O/L
500 mL = 10.6 mg H2O/L
750 mL = 12.9 mg H2O/L
1,000 mL = 14.3 mg H2O/L
1,250 mL = 14.9 mg H2O/L
250 mL = 8.7 mg H2O/L
500 mL = 11.1 mg H2O/L
750 mL = 12.9 mg H2O/L
1,000 mL = 14.7 mg H2O/L
1,250 mL = 15.7 mg H2O/L
Resistance
Before
Use
Flow at 0.5 L/sec = 0.6 cmH
2O
Flow at 1.0 L/sec = 1.6 cmH
2O
Flow at 1.5 L/sec = 3.3 cmH
2O
Flow at 0.5 L/sec = 1.3 cmH
2O
Flow at 1.0 L/sec = 3.3 cmH
2O
Flow at 1.5 L/sec = 6.1 cmH
2O
Resistance
After
Use
Flow at 0.5 L/sec = 0.7 cmH2O
Flow at 1.0 L/sec = 2.0 cmH
2O
Flow at 1.5 L/sec = 4.5 cmH
2O
Flow at 0.5 L/sec = 1.5 cmH
2O
Flow at 1.0 L/sec = 4.0 cmH
2O
Flow at 1.5 L/sec = 6.0 cmH
2O
003020
Air
Life
™
Nonfiltered BHME
003021
Air
Life
™
Filtered BHME
Filtration
Efficiency
N/A
250 mL = 9.1 mg H2O/L
250 mL = 9.1 mg H2O/L
250 mL = 9.1 mg H2O/L
Recommended
Change Out
24 hours 24 hours
Connections
15 mm ISO male connector for Y piece
15/22 mm ISO female/male connector for ET tube,
closed suctioning system, etc.
Gas Leakage 0 cc/min at 7 kPa (71 cmH
2O) 0 cc/min at 7 kPa (71 cmH2O)
Compliance 0.68 mL/kPa at 7 kPa (71 cmH2O) 0.68 mL/kPa at 7 kPa (71 cmH2O)
MR Conditional
Field strength = 3 – Tesla or less
Spatial gradient = 720 Gauss/cm or less
Radio frequency (RF) field = 128 MHz
Specific absorption rate = 2.9 W/kg
Packaging 25 units per case 25 units per case
AirLife
™
BHME Humidifiers
Optional filter helps prevent
respiratory complications
and reduces chance of
cross-contamination
Aerosol mode
Gate open
Lever down
Aerosol mode
HME mode
Lever up
Gate closed
HME mode
Bypass lever position and clear visual indicators ensure gate engagement
AirLife
™
Bypass HME
AirLife
™
bypass HME (BHME) technology delivers aerosolized medication treatments without breaking the ventilator circuit. Easy one-handed
operation allows bypass engagement for simple medication delivery setup. The bypass lever position and clear visual indicators help identify
gate engagement. The filtered BHME helps prevent respiratory complications and reduces the risk of cross-contamination.
Cat. No. Description Qty.
003020 Air
Life
™
nonfiltered bypass HME 25/cs
003021 Air
Life
™
filtered bypass HME 25/cs